1. Analytical Precision
Analytical susceptibility study was performed with known values and recombinant antigens commercially available for SARS Cov-2 virus.
The analytical susceptibility chart obtained as a result of the tests performed with the specified antigens is given in the table.
Nucleoprotein: 100 picograms
2. Inactive Virus Susceptibility
In the dilution study performed with 109 pfu/ml inactivated SARS Cov-2 virus, the sensitivity of 1/20,000 dilution and the inactivated virus was 40,000 pfu/ml.
3. Sars Cov 2 Covid 19 Antigen Rapid Test Clinical Performance
Samples from 93 RT-PCR-identified positive patients were processed with this assay, and 90 samples were detected as positive; On the other hand, 327 samples were detected as negative in samples taken from 3222 RT-PCR-identified negative patients.
Positive Result | Negative Result | Invalid Result |
---|
If both the (C) and (T) lines appear within 15 minutes, the test result is positive and valid. | If the test area (T) has no line and the control area shows a colored line, the result is negative and valid. | If a colored line does not appear in the control region (C), the test result is invalid. The sample should be retested using a new test cassette. |