Catalog NO; LC001.V1
DNA/RNA Hunter Device, DNA/RNA Hunter Device Charger and User Manual
Battery specifications: 3 pieces with a capacity of 3.7 VDC @ 1500 mAh, a total of 11.1 VDC @ 1500 mAh, Energy equivalent of 16.65 W.h 60 kjoules.
Power and operating time: The average power drawn in the total system has been measured at around 10 watts, and in terms of battery capacity; Considering that there may be changes due to environmental and climatic conditions for a fully charged battery, it is expected that the average operating time will be around 90 minutes.
Heater temperature limit: It is designed to reach the maximum temperature of 110 degrees in 2 minutes under normal climatic conditions or at room temperature, and the mechanical and electronic materials used are selected to withstand 150 degrees for a short time (max 2 minutes) and 120 degrees for an indefinite period.
Top cover temperature: Since it is one of the materials with a very high temperature resistance (250 degrees) between the top cover heater and the cover so that the top cover temperature is kept low, and the thermal conductivity coefficient is very low (5. 10E-4 Cal / cm.s. degree ) 4 mm thick Teflon plate was used. As a result of the calculations and tests, it was observed that while the internal temperature was 80 degrees, the upper lid temperature remained between 30-35 degrees.
Cooling time: The active cooling time is 30 seconds while the fans are running, and the passive cooling time is more than 15 minutes due to thermal insulation. Warm-up and cool-down time: It has been observed that the total warm-up and cool-down time is less than 2 minutes between 20 degrees and 60 degrees.
Display features: 3.5 inch Resistive touch panel 320 x 480 pixel TFT full color graphic display.
Optical evaluation: Full color digital color sensor is used and the measured color values are compared with reference values and the result of the preliminary evaluation is displayed on the screen by the system, and the actual colors are also reflected on the screen.
Test program features: Test recipes can be accessed only by giving the transaction code, thanks to the prescription and recipe memory processes for 15 different processes. For each test process, three separate temperature set values and warm-up times are determined in the recipes, and if necessary, the option of temporary or permanent correction in the recipes is offered to the user. Transaction results can be exported to sd cards and then moved to the host. The test results, which will be matched over the 10-digit identification codes for the samples, will be recorded within the matching codes and stored on the sd card.
Total weight and dimensions: Total weight is 430 gr and external dimensions are 70 x 175 x 80 mm.
With the Biodmes DNA/RNA Hunter device, it can give fast and precise results. DNA/RNA Hunter device provides amplification of DNA or RNA by isothermal amplification method. At the end of the amplification period, the results can be read with the naked eye, and the results can be evaluated with the color reader system of the DNA/RNA Hunter device. The device can work for the desired time and at a constant temperature, and also offers the opportunity to adjust the temperature and time.